New ultrasound transducers should include an integrated quality control check that runs every time the probe is turned on, according to final guidance issued by the U.S. Food and Drug Administration (FDA) on August 22.
The FDA plans to incorporate the new transducer check recommendation into the agency's 510(k) compliance policy that covers marketing clearance of diagnostic ultrasound systems and transducers. The new policy -- as well as clarifications to guidelines on when the FDA will require modified probes to receive a new 510(k) application -- was explained in a webinar by Shahram Vaezy, PhD, of the FDA's Center for Devices and Radiological Health.
"This guidance supersedes the FDA's 2008 guidance ... and describes the types of modifications to a diagnostic ultrasound device for which the FDA does not intend to enforce the requirement for a premarket notification," Vaezy said. "[It also] includes a new transducer element check that applies to all ultrasound devices."
Transducer element check
The FDA is recommending that manufacturers integrate into transducers for which they are filing new 510(k) applications a way to test probe performance each time it is activated.
"Each device should include some level of testing," the agency wrote. "This integrated test feature would also generate a report on the performance of the probe under test for documentation ... [and] should also be available to the operators to initiate anytime when a particular probe is suspected of failure."
Developing transducer check features will take time, Vaezy acknowledged.
"The FDA is open to discussions with manufacturers about how to implement transducer checks into devices," he said.
Compliance policy
The FDA clarified that it will not require a new 510(k) for modified ultrasound and transducer devices that have already received an initial 510(k) clearance when all of the following apply:
- The intended use of the modified device is not changed.
- The device is not a reusable device subject to the requirement for the submission of reprocessing labeling and validation data.
- The modes of operation for the modified device are well-established.
- The modifications do not lead to acoustic outputs that exceed the recommended maximum acoustic output levels.
- The modifications do not result in a range of ultrasound interrogation parameters outside a well-known range.
- The modifications do not use novel mechanical or thermal effects for imaging or measurements.
- The measurements and analyses are clearly described, and the user can adjust the associated control parameters.
- Transducer element check is performed.
- Transducer surface temperature falls within a well-defined range.
- Appropriate transducer covers are recommended to users.
During the webinar, Vaezy listed some examples of possible compliance policy applications that would require a new 510(k):
- Adding continuous-wave and pulsed-wave Doppler methods to the device
- Adding an algorithm that measures the volume of an organ based on established image segmentation and volume calculation methods
- Adding a new transducer with similar indications for use and similar acoustic output
- Adding a B-mode noise reduction filter for general imaging use to a system
"The revised final guidance enables manufacturers with an established track record of ultrasound device development, via submission of 510(k)s for their original devices, to add new safety features and make certain modifications to address clinical needs without [having to] submit another 510(k)," he said.

![Examples of ultrasound findings and techniques. (A) Images in a 39-year-old male patient with a mass in the left thigh. The mass is heterogeneous on the B-mode US image (compared with the patient in D) and showed increased microvascularity (superb microvascular imaging [SMI]) and shear-wave elastography (SWE) values. Undifferentiated pleomorphic sarcoma was diagnosed at biopsy (with pleomorphic rhabdomyosarcoma in surgical specimen). (B) Images in an 18-year-old male patient with a mass in the left leg. The mass is hypoechoic on the B-mode image, with no other findings suggestive of malignancy. The lesion is in contact with the cortex of the tibia, which is slightly irregular. CT revealed a doubtful anteromedial tibial erosion. The microvascular study demonstrated high vascularization, suggestive of malignancy. Periosteal Ewing sarcoma was diagnosed with both histologic and immunohistochemical confirmation. (C) Images in a 69-year-old female patient with a lump growing on the outside of the left leg. Multiple SWE examinations were performed (please note the high values obtained in the measurements, whereas the color map highlights the stiffness relative to adjacent tissues). SMI showed areas of increased vascularization to target for sampling. Undifferentiated spindle cell sarcoma was diagnosed at biopsy, with residual leiomyosarcoma in the surgical specimen after neoadjuvant therapy. (D) Images in a 56-year-old female patient with a mass in the right thigh. The mass is heterogeneous at both B-mode ultrasound (similar to patient A) and MRI (coronal T2-weighted spectral attenuated inversion recovery [SPAIR]; T1-weighted pre-contrast and postcontrast imaging), which even shows uptake after the administration of paramagnetic contrast material, which is traditionally suggestive of malignancy. Low values at SMI and elastography are suggestive of benignity. Spindle cell lipoma was diagnosed at biopsy, with atypical spindle cell lipomatous tumor in the surgical specimen.](https://img.auntminnie.com/mindful/smg/workspaces/default/uploads/2026/08/images-radiol250278fig2.APCFLSvX6p.jpg?auto=format%2Ccompress&fit=crop&h=100&q=70&w=100)





![Examples of ultrasound findings and techniques. (A) Images in a 39-year-old male patient with a mass in the left thigh. The mass is heterogeneous on the B-mode US image (compared with the patient in D) and showed increased microvascularity (superb microvascular imaging [SMI]) and shear-wave elastography (SWE) values. Undifferentiated pleomorphic sarcoma was diagnosed at biopsy (with pleomorphic rhabdomyosarcoma in surgical specimen). (B) Images in an 18-year-old male patient with a mass in the left leg. The mass is hypoechoic on the B-mode image, with no other findings suggestive of malignancy. The lesion is in contact with the cortex of the tibia, which is slightly irregular. CT revealed a doubtful anteromedial tibial erosion. The microvascular study demonstrated high vascularization, suggestive of malignancy. Periosteal Ewing sarcoma was diagnosed with both histologic and immunohistochemical confirmation. (C) Images in a 69-year-old female patient with a lump growing on the outside of the left leg. Multiple SWE examinations were performed (please note the high values obtained in the measurements, whereas the color map highlights the stiffness relative to adjacent tissues). SMI showed areas of increased vascularization to target for sampling. Undifferentiated spindle cell sarcoma was diagnosed at biopsy, with residual leiomyosarcoma in the surgical specimen after neoadjuvant therapy. (D) Images in a 56-year-old female patient with a mass in the right thigh. The mass is heterogeneous at both B-mode ultrasound (similar to patient A) and MRI (coronal T2-weighted spectral attenuated inversion recovery [SPAIR]; T1-weighted pre-contrast and postcontrast imaging), which even shows uptake after the administration of paramagnetic contrast material, which is traditionally suggestive of malignancy. Low values at SMI and elastography are suggestive of benignity. Spindle cell lipoma was diagnosed at biopsy, with atypical spindle cell lipomatous tumor in the surgical specimen.](https://img.auntminnie.com/mindful/smg/workspaces/default/uploads/2026/08/images-radiol250278fig2.APCFLSvX6p.jpg?auto=format%2Ccompress&fit=crop&h=112&q=70&w=112)