Tuesday, November 30 | 8:35 a.m.-8:45 a.m. | VV31-02 | Room E353A
To reduce CT radiation exposure, radiologists have gotten used to adjusting kV based on body mass index and the diagnostic task at hand. But how about doing it quantitatively and automatically?Researchers from the Mayo Clinic in Rochester, MN, developed a novel tool that permits automatic adaptation of kV as a function of patient size and diagnostic task, as well as attenuation levels on the CT scout image in abdominal and thoracic imaging.
Lifeng Yu, PhD, and colleagues performed two experimental studies, one using four thoracic phantoms and the other using seven abdominal phantoms, all packed with iodine samples to measure contrast-to-noise ratio. The goal was to assess the achievable radiation doses and examine workflow patterns.
For the four thoracic phantoms, the optimal kVs were 80, 80, 100, and 100, with corresponding relative doses of 37.6%, 44.4%, 60.8%, and 89.5%, compared to standard 120-kV imaging. For the seven thoracic phantoms, the most dose-efficient kV settings ranged from 80 to 140 kV, with corresponding radiation doses relative to 120-kV scans of the same phantom at 57.8% to 100%. Depending on patient size, optimal kV settings suggest dose reductions of up to 62.4%, Yu said.














![Axial images from unenhanced calcium score cardiac CT (left) and curved planar reformation images from CT angiography (right) show that higher long-term exposure to air pollution is associated with greater coronary artery calcium and more obstructive coronary artery disease (CAD). Top row: Images in a 68-year-old male patient with higher 10-year mean ambient air pollution exposure (7.9 μg/m3 for particulate matter measuring ≤2.5 μm in diameter [PM2.5] and 17.4 parts per billion [ppb] for NO2) with extensive CAD (coronary artery calcium score [CACS] >1,000 and obstructive CAD [≥70% diameter stenosis]). Bottom row: Images in a 57-year-old female patient with lower 10-year mean ambient air pollution exposure (6.3 μg/m3 for PM2.5 and 4.6 ppb for NO2) with no CAD (CACS = 0 and no obstructive stenosis).](https://img.auntminnie.com/mindful/smg/workspaces/default/uploads/2026/06/hanneman.r6SMLzkezo.png?auto=format%2Ccompress&fit=crop&h=112&q=70&w=112)