By Donna Domino, AuntMinnie.com staff writer

September 18, 2009 -- SACRAMENTO, CA - A California radiologic technologist accused of delivering a major radiation overdose to a 23-month-old boy faced testimony this week on her role in the incident. Witnesses who spoke at a hearing on revoking the technologist's license said they still don't understand how the boy received 151 CT scans in just over an hour.

In January 2008, 23-month-old Jacoby Roth fell out of bed, and when he could hardly move his head his worried parents took him to the emergency room at Mad River Community Hospital in Arcata, a small town 290 miles north of San Francisco.

There, a doctor ordered x-rays and CT scans to check for damage to the child's cervical spine. He was taken to the scanning room, where radiologic technologist Raven Knickerbocker performed CT scans at C-spine levels C1 through C4 in the same section of the midmaxillary sinuses, midclivus, and posterior fossa.

Over the next 68 minutes, the toddler was exposed to 151 scans.

Radiation overdose

The incident was recalled this week in two days of hearings held by the California Department of Public Health (CDPH) on whether to revoke Knickerbocker's radiologic technologist license.

In testimony before an administrative law judge, the boy's father, who during the scans was standing at the foot of the CT table to calm his son, recalled his growing concern as the scanning stretched on. "I said, 'Stop this!' " Padre Roth recalled, noting that Knickerbocker finally stopped the scan only after he became angry.

Within a few hours, the child developed a bright red ring around his head from the massive overdose of radiation. Photographs of the left side of the boy's face show a clear line extending from the infraorbital ridge backward through the ear and nape of the neck; a similar line extends from the infraorbital ridge through the ear on the right side.

Jacoby Roth several hours after receiving 151 CT scans in a 68-minute period. Photo courtesy of Roth family attorney Don Stockett.
In off-the-record comments, some state officials called it the worst case of radiation overdose of a child in the U.S.

The boy's mother, Carrie Roth, recalled questioning the necessity of the scans, and was concerned about radiation because her son had already had a CT scan when he was younger. "Is this safe?" she remembered asking doctors when she was told her son needed yet another CT scan because he had moved during the scans done by Knickerbocker.

Padre Roth asked how that could be, noting that his son was in a papoose board restraint and was asleep for most of the first scanning session.

"I had a bad feeling," he testified. "I knew something was wrong; there's no way a CT scan is going to take an hour and 20 minutes. I should have stopped it sooner."

"I started crying," the boy's mother recalled. "What happened the first time?" she asked the doctor, referring to the first lengthy scan.

Carrie Roth described seeing Knickerbocker "push the [scanner] button many times, about every 30 seconds."

Scanner records show an average interval of 25 seconds between the 151 scans, which started at 8:29 a.m. on January 23, 2008, and ended at 9:37 a.m. The CT scanner used was a single-slice Picker PQ-5000 manufactured in 1998, which needed about 25 seconds for the tubes to cool between scans. It was replaced a month after the incident, but hospital officials said the replacement had been planned long before and was unrelated.

A second CT scan by radiologic technologist Susan Sampson took only two minutes and included 25 axial slices. Sampson remembered being "a little shocked" when she heard that Knickerbocker had done 151 scans in the same area at the base of the boy's skull. "That's a lot more time than it usually takes for an exam," she testified, estimating it should have taken less than 10 minutes.

Knickerbocker showed little emotion during the hearing, but became teary when a former supervisor said she had been a good employee before the incident. She did not testify during the hearing.

Bruce Fleck, the hospital's former radiology manager, testified that Knickerbocker subsequently gave many explanations for the incident, such as the boy's parents distracted her, the scanning table wouldn't move incrementally, and the boy's father was leaning on the table. But he noted that an experienced operator like Knickerbocker should have known to stop after a couple of images if the scanner wasn't operating properly.

When asked if the scanner could somehow take images automatically, he replied that the images showed the machine was in manual axial mode. "She had to hit the button each time," Fleck observed.

He also noted that Sampson had no problems when she used the machine. It was checked later and no malfunctions were found.

Knickerbocker could not provide a "valid explanation" of why she took 151 images of the same location, Fleck said. "I think it was just a rogue act of insanity," he told a stunned court.

“I think it was just a rogue act of insanity.”
— Bruce Fleck, former radiology
    manager, Mad River
    Community Hospital

Dave Laumann, the head technologist at the time, said Knickerbocker told him she didn't know 151 scans had been taken. "It just seemed to me that she wasn't dealing with it," he testified.

Fleck was so upset by what he called Knickerbocker's "egregious ethical breach" that he wrote to the American Registry of Radiologic Technologists (AART) about the incident. Many of the hospital's radiology staff said Knickerbocker violated ALARA (as low as reasonably achievable) standards by doing such a lengthy scan.

A report by the hospital's medical physicist calculated that the boy's absorbed radiation dose was 2.8 Gy (2,800 mSv) and possibly as high as 11 Gy (11,000 mSv). The dose the boy received compares with a range of 1.5 to 4.0 mSv for a normal pediatric CT study of the entire spine, according to pediatric imaging experts.

Using relevant material from the article "Estimated Risks of Radiation-Induced Fatal Cancer from Pediatric CT," published in the American Journal of Roentgenology (February 2001, Vol. 176:2, pp. 289-296), a report by the hospital's medical physicist concluded the child had a lifetime increased risk of a fatal cancer of 39%.

State investigators who interviewed Knickerbocker said she told them that she only pushed the scanner button two or three times, and could not explain why the machine took so many scans.

Mindy Malone, a CDPH health physicist who investigated the incident, said Knickerbocker called for help during the scanning session but never got any. But Laumann said he came into the room shortly after the scanning began to check on the procedure, contradicting Knickerbocker's assertions.

Dr. Wesley Root, the hospital's head of radiology, said a scanner check showed that Knickerbocker had plotted only the beginning slice for the exam, meaning the machine would automatically stop after one slice, Malone testified.

"It did one scan at a time," Malone testified. "She had to manually make the exposure every time."

Root also noted that the red light over the scanner's gantry flashes each time a scan is done, casting doubt on Knickerbocker's statements that she wasn't aware that 151 scans had been acquired.

Suspended license

Mad River fired Knickerbocker two weeks after the incident; the CDPH suspended her license nine months later. The CDPH determined that "operator error" by Knickerbocker caused the radiation overdose and fined Mad River $25,000, noting that the hospital failed to follow its policies and procedures for radiation safety. Knickerbocker gave inconsistent explanations about the incident, state officials found.

The Roth family's attorney, Don Stockett of Folsom, CA, has filed a lawsuit against the hospital and Knickerbocker, claiming negligence and medical battery. Stockett said he sent a blood sample from Jacoby Roth for analysis to Dr. David Lloyd, a DNA specialist and principal investigator of the Molecular Design Group of the School of Biochemistry and Immunology at Trinity College in Dublin, Ireland. The analysis found that Roth sustained substantial chromosomal damage, Stockett said.

Knickerbocker had been expected to testify at the Sacramento hearing, but because the testimony ran long, the hearing has been continued until a future date yet to be scheduled.

Padre Roth said Knickerbocker should lose her license for good. "We're still worried about the long-term effects on our son," he said. "She should never be allowed near anyone's children again."

By Donna Domino
AuntMinnie.com contributing writer
September 18, 2009

California hospital fined $25,000 for pediatric CT radiation overdose, March 24, 2009

Parents sue California hospital over CT radiation dose, November 20, 2008

Copyright © 2009 AuntMinnie.com


Last Updated hh 4/14/2011 10:22:51 AM

43 comments so far ...
9/18/2009 10:49:42 AM
IR or die
Nice article.... keep up the good work auntie

9/18/2009 10:52:05 AM
X-Ray Eyes
Somehow I missed this when it first happened, and I nearly cried reading this article. Then I wanted to beat the tech within and inch of her life and then scan HER head repeatedly! This is a clear case of someone who should not be able to EVER touch an x-ray machine again! I cannot understand why she did not get help if she was having an issue with the machine/scan (why else would you keep imaging the same level 151 times?) unless you are a completely demented individual, who should be put behind bars. And shame on the doctors and staff for leaving her alone for over one hour to do a head exam! It doesn't matter that it was a single-slice machine, I trained on one and it did not take anywhere near that long for and adult head, let alone a kids head. They should have been in there assessing the situation and deciding to reschedule or forgo the scan altogether.
I just see so much wrong with what happened and, as a father with small kids, it just makes me sick knowing that there are some "techs" out there that have no business representing my profession!
I will now climb off my soap box.

9/18/2009 11:15:14 AM
Adeelmd
You don't do over 100 scans by mistake. this sounds like it was done on purpose. Why?

9/18/2009 12:20:49 PM
TarverForPresident
Quote from
ORIGINAL: Adeelmd

You don't do over 100 scans by mistake. this sounds like it was done on purpose. Why?

 
Agreed. There's alot more to this case than we know so far.

9/18/2009 12:31:25 PM
gbnoethen
There is alot of important information left out of this article, such as experience level of the technologist and was she a registered CT technologist? Also, what was her experience level on this machine. It sounds to me with the information given in the article that the first scan had motion and had to be repeated. It also sounds like she was trying to get the image ot the same level as the previous scan that had motion, without running a completely new scan. Either she couldn't figure it out and repeatedly scanned the same area trying to get the correct level. or couldn't figure out how to set the scan up correctly.
Everything in this article screams of inexperience and unfamiliarity with the machine and modality. I agree, she should of stopped the scanner and taken the patient back tthe ER and had another technologist called. It sounds like that is what she did..... later.... By not understanding CT and the high does associated with it she repeatedly scanned the patient unaware of the dose she was delievering.
Even in standard film radiography, after 3 repeats you should be out getting help or another technologist opinion. Shame on the hospital, department manager, lead CT technologist for allowing this technologist to run a scanner alone.
Just a guess, this was a second shift RT that was trying to get the job done without calling the CT call tech in.
This folks is why we have registries and proficency exams. I hope tools are put in place to keep this from happening again.
There are alot of fine imaging technologist out there, this in not a representation of the people in our field. I hope it was inexperience and nothing else.
I have 14 years in this profession and feel so bad for the child and parents.

Jerry N RT (R) CT