Ultrasound best in margin assessment for breast surgery

2021 06 16 20 29 9882 Doctor Ultrasound Keyboard 400

Diagnostic ultrasound outperforms other imaging methods when it comes to assessing margins of cancers being excised during breast-conserving surgery, according to research published January 4 in Academic Radiology.

A team led by Dr. Irina Manhoobi from Aarhus University Hospital in Denmark found that ultrasound had the highest pooled sensitivity, specificity, and area under the curve (AUC) for margin assessment compared to radiography, digital breast tomosynthesis (DBT), and micro-CT.

"Achieving adequate resection margins in breast conserving surgery is challenging and often demands more than one surgical procedure," Manhoobi and colleagues wrote.

Breast-conserving surgery and whole-breast irradiation are the standard treatments for early-stage breast cancer. One challenge facing surgeons, however, is that breast-conserving surgery involves removing the tumor with a sufficient margin while also preserving the physical appearance of the breast.

Not much is known when it comes to the pooled diagnostic performance of imaging methods during surgery for assessing tumor margins, as well as their impact on repeat surgery, according to the researchers.

Manhoobi et al wanted to look at the sensitivity and specificity of conventional and emerging imaging methods for margin assessment during breast-conserving surgery. They also wanted to find out if these methods contribute to a reduced repeat surgery rate.

They looked at 22 research articles from the U.S., Europe, and Asia with 29 radiological imaging methods for meta-analysis, with the articles using either ultrasound, radiography, DBT, or micro-CT. The team found that ultrasound achieved the highest marks, despite only one study showing its use in assessing margins during surgery. The rest of the included studies showed that ultrasound was used for preoperatively marking the breast tumor.

Performance of imaging modalities for breast‑conserving surgery
  Radiography DBT Micro-CT Ultrasound
Sensitivity 52% 67% 68% 72%
Specificity 77% 76% 69% 78%
AUC 60% 76% 72% 80%

Despite ultrasound's high marks, the researchers stated it is "not realistic" that ultrasound of specimens will be implemented in surgical workflows.

"It is too resource demanding for a breast radiologist to attend the operating theater and perform the intraoperative ultrasound of the specimen, and it takes time for a breast surgeon to achieve the same level of experience with ultrasound as a breast radiologist," the authors wrote.

Ultrasound's higher level of performance could be linked to operator dependability, they noted. Experienced breast radiologists performed ultrasound scans in half of the included studies. In addition, ultrasound is limited by difficulties of detecting ductal carcinoma in situ (DCIS), and its use risks flattening the specimen when the transducer meets the specimen's surface, leading to false-positive findings.

"The radiological methods for margin assessment need further improvement to provide reliable guidance in the clinical workflow and to prevent repeat surgeries for early-stage breast cancer patients," the researchers concluded.

Page 1 of 569
Next Page