Timing is everything in contrast-enhanced CT
Article Thumbnail ImageSeptember 28, 2006 -- When clinicians need to image the abdominal aorta and the lower extremity arterial tree in patients with peripheral arterial occlusion disease (PAOD), matching the injection of contrast to when CT scanning begins in those regions is key.
Member Sign In:
  MemberID or E-mail Address:
  Password:
(?)

New AuntMinnie.com Members

Becoming a Member is FREE!

  • Real-time radiology-specific news
  • Case of the Day and Teaching Files
  • Focused topics digital communities
  • Lively, discussion groups
  • Medical imaging bookstore
  • SalaryScan
  • Job Boards
  • OnLine CMe/CE
  • Bookstore, market reports, more …
  • Conference Calendar
  • User-controlled eNewsletters
  • … registration is FREE and easy!
  • Dr. Dominik Fleischmann, associate professor of radiology at Stanford University in California, has developed protocols for optimal wait times between contrast injection and the start of CT imaging based on how fast the bolus reaches the abdominal aorta or arterial tree. Fleischmann presented his findings at Stanford's 2006 International Symposium on Multidetector-Row CT in San Francisco.

    Patients with cardiovascular disease and limited cardiac output often have problems with arterial enhancement following contrast injection. The condition may lead to inadequate opacification of the abdominal aorta. "We know that the bolus transit down the peripheral territory can be substantially delayed in patients with occlusive disease," Fleischmann told attendees.

    Of course, outrunning the bolus is a real problem with today's fast multidetector scanners. If imaging starts when contrast reaches the aorta, CT acquisition -- in Fleischmann's words -- "may in fact be faster than the bolus transit speed down the legs and feet." The question is whether clinicians can scan faster than the bolus travel time to the peripheral arterial tree.

    Travel time

    To find the answer, Fleischmann and his colleagues calculated the timing on 20 patients with PAOD by injecting a test bolus to measure time attenuation curves in the abdominal aorta. The results showed that the bolus took approximately 10 seconds to travel from the aorta, but the timing varied per patient, depending on the degree of PAOD. In some patients, travel time took only four seconds (177 mm/sec), while it took as long as 24 seconds (33 mm/sec) in other patients.

    "What you can extrapolate from that is that it takes up to 48 seconds to fill the entire peripheral arterial tree, and you can inject five seconds shorter than your scan time," Fleischmann said.

    From that starting point, clinicians can adjust injection and scan times accordingly. With an acquisition speed of 33 mm/sec, the risk of outrunning the bolus is very low. "If you have a speed of 6 mm/sec -- which you can easily achieve with a 16-row scanner -- you are probably scanning faster than the bolus moves down the aorta in about one-third of your patients," Fleischmann added. Using a 64-row CT with an acquisition speed of 19 mm/sec, a clinician would scan faster than the bolus travel time to the arterial tree in approximately 50% of the patients.

    A head start

    To solve the problem, Fleischmann recommends increasing the contrast media injection rate for the first five seconds of the injection, or increasing the scanning delay relative to the arrival of contrast media in the aorta, allowing more time for the contrast to enhance the aorta.

    "You need to give the bolus a head start," Fleischmann explained. "Say the scan time is only 25 seconds. You wait until the contrast arrives in the aorta, then wait another 15 seconds and then catch up so that overall you are ready in 40 seconds when the contrast arrives from the aorta. This also very reliably gives you very bright arterial opacification, despite the presence of substantial arterial occlusive disease."

    Integrated 64-channel peripheral CTA acquisition
    and injection protocol
    Acquisition 64-row by 0.6 mm (channels by channel width)
    Pitch, table speed Variable, depending on volume coverage
    Scan time Fixed to 40 seconds in all patients
    Injection duration Fixed to 35 seconds in all patients
    Scanning delay Aortic contrast transit time + 2 seconds
    Contrast medium 370 mgI/mL
    Biphasic injections Maximum flow rate for first five seconds of injection, continued with 80% of initial flow rate for 30 seconds
    Source: Dr. Dominik Fleischmann, Stanford University

    When using a 64-slice CT scanner, Fleischmann said the protocol at Stanford is to slow down the system and scan for 40 seconds, regardless of the size of the patient. With a variable pitch setting and by alternating the CT tube current, this reduces radiation to the patient and provides reliable images across a wide range of patients.

    By Wayne Forrest
    AuntMinnie.com staff writer
    September 28, 2005

    Related Reading

    New contrast injectors go with the digital flow, September 28, 2006

    Grown-up decisions: When is MDCT the best choice in children? September 30, 2005

    Part II: Palliative steps prevent contrast-induced nephropathy, September 20, 2006

    Part I: Identifying patients at risk of contrast-induced nephropathy, September 13, 2005

    Low-dose CT matches x-ray for pediatric sinus imaging, June 9, 2005

    Copyright © 2006 AuntMinnie.com


    || About || Advertising || Bookstore || Breast MRI || Career Center || Case of the Day || Communities || Conferences || Contact Us || ECR News 2013 || Education || Equipment Classifieds || Facebook || Forums || Home || Links || Marketplace || Mobile || Molecular Breast Imaging || New Installations || News in Brief || People in the News || Privacy Policy || RSNA News 2012 || Reference || Salary Survey Results || Trends in Radiology || Twitter || Vendor Connect || Webinars || XML/RSS ||

    Copyright © 2013 AuntMinnie.com. All Rights Reserved.